SELECTION OF INTERNAL STANDARDS AT THE ANALYSIS OF BIOLOGICAL OBJECTS IN THE MASS-SPECTROMETER WITH INDUCTIVLY COUPLED PLASMA (İCP-MS)

Authors

  • Nurubeyli T.K. Institute of Physics of the Azerbaijan National Academy of SciencesG.Javid ave. 131, AZ-1143, Baku, AZERBAIJAN
  • Hasanova S.I. Institute of Physics of the Azerbaijan National Academy of SciencesG.Javid ave. 131, AZ-1143, Baku, AZERBAIJAN

Keywords:

Matrix Effect, Biological Liquid, Internal Standard, Atomic Weight, Ionization Potentials, Mass-Spectrometric Analysis

Abstract

In work it was established that in a "standard" mode of a mass spectrometer when accounting all possible matrix effects, after a dilution of biological liquids the exact definition of all studied set of elements of one solution is possible only in case of application of several internal standards (IS) which are picked up proceeding from a proximity to their ionization potential (IP) to IP of anolytes. It is also shown, the nature of the element acting as the internal standard is not basic, and so, reduction of set of the applied VS to unique is possible. The last approach is especially simple performed by that it is important for clinical practice.

Additional Files

Published

07-10-2023

How to Cite

Nurubeyli T.K., & Hasanova S.I. (2023). SELECTION OF INTERNAL STANDARDS AT THE ANALYSIS OF BIOLOGICAL OBJECTS IN THE MASS-SPECTROMETER WITH INDUCTIVLY COUPLED PLASMA (İCP-MS). International Educational Applied Scientific Research Journal, 3(5). Retrieved from https://ieasrj.com/journals/index.php/ieasrj/article/view/101